Page 1 - Series 50 IP-2 (M1353A)

Series 50 Fetal MonitorsSeries 50 A (M1351A)Series 50 IP-2 (M1353A)SERVICE AND INSTALLATION GUIDEM1353-9000JPrinted in Germany April 2002

Page 2

x Contents

Page 3 - Printing History

Toco Parameter Test90 Chapter 8 Troubleshooting Flowcharts Toco Parameter Test Figure 8-19 Troubleshooting: Toco Parameter TestCPU Boa

Page 4

Maternal NIBP with the Dinamap 1846/8100 MonitorChapter 8 Troubleshooting Flowcharts 91Maternal NIBP with the Dinamap 1846/8100 Monitor

Page 5 - Contents

Maternal NIBP with the COLIN Model BP-8800 Monitor92 Chapter 8 Troubleshooting Flowcharts Maternal NIBP with the COLIN Model BP-8800 Monitor

Page 6

Fetal Pulse Oximetry with Nellcor N-400 or Compatible MonitorChapter 8 Troubleshooting Flowcharts 93Fetal Pulse Oximetry with Nellcor N-400 or

Page 7 - 9. Preventive Maintenance

Paper Sensing Test94 Chapter 8 Troubleshooting Flowcharts Paper Sensing Test Figure 8-23 Troubleshooting: Recorder Paper SensingCaut

Page 8

Chapter 9 Preventive Maintenance, Care and Cleaning 959Preventive Maintenance,Care and CleaningIntroductionThis chapter contains maintenance an

Page 9 - Contents ix

Regular Maintenance96 Chapter 9 Preventive Maintenance, Care and Cleaning Regular MaintenanceThis section details the periodic maintenance recom

Page 10

Accessory TestingChapter 9 Preventive Maintenance, Care and Cleaning 97Accessory TestingTesting Toco TransducersVisual check Ensure that there

Page 11 - General Information

Accessory Testing98 Chapter 9 Preventive Maintenance, Care and Cleaning Testing Ultrasound TransducersVisual check Ensure that there are no crac

Page 12 - Protective Earth Terminal

Accessory TestingChapter 9 Preventive Maintenance, Care and Cleaning 996. Holding the transducer in one hand, move a flat-bottomed pencil or si

Page 13 - Initial Inspection

Chapter 1 General Information 11General InformationIntroductionThis guide tells you how to install, service, and repair an Series 50 A (M1351A)

Page 14 - Overview

Safety Testing100 Chapter 9 Preventive Maintenance, Care and Cleaning The Toco display shows 0.TOCO int, time, date and paper speed are printed

Page 15 - Monitors

Safety TestingChapter 9 Preventive Maintenance, Care and Cleaning 101documentation included with those products.When performing safety tests, y

Page 16

Safety Testing102 Chapter 9 Preventive Maintenance, Care and Cleaning When to Perform Safety Tests Table 9-2M1351A and M1353A: When to perform s

Page 17

Safety TestingChapter 9 Preventive Maintenance, Care and Cleaning 103Test and Inspection MatrixKey to Table: P = Pass and F = Fail, X = test re

Page 18

Safety Testing104 Chapter 9 Preventive Maintenance, Care and Cleaning Safety TestsS(2): ProtectiveEarth TestTest to perform: The protective eart

Page 19

Safety TestingChapter 9 Preventive Maintenance, Care and Cleaning 105Instrument SafetyTestYou must perform the instrument safety test every tim

Page 20 - TraceVue. English only

Safety Testing106 Chapter 9 Preventive Maintenance, Care and Cleaning

Page 21 - Accessories Option Model

Chapter 10 Peripherals 10710PeripheralsFitting the Combined Interface ModuleTo fit the Combined Interface Module: 1. Remove the s

Page 22

Connecting Peripheral Devices108 Chapter 10 Peripherals Connecting Peripheral DevicesIf the monitor has the Combined Interface Module, you can c

Page 23 - Technical Specifications

RS232 Serial InterfaceChapter 10 Peripherals 109RS232 Serial Interface Figure 10-3RS232 Interface Pin ConnectionsThe following tab

Page 24 - Inputs M1351A Single

Conventions and Symbols Used in this Guide2 Chapter 1 General Information Conventions and Symbols Used in this GuideThis guide uses the followin

Page 25 - DECG and MECG

RS232 Serial Interface110 Chapter 10 Peripherals Barcode Reader After installing the Barcode Reader, you must check that it is connected correct

Page 26 - 1846/8100

RS232 Serial InterfaceChapter 10 Peripherals 111NIBP Monitor One of the following devices can be connected to an M1351A or M1353A to record mat

Page 27

RS232 Serial Interface112 Chapter 10 Peripherals To connect a COLIN Model BP-8800 Monitor to the M1351A or M1353A:1. Use the interface cable sup

Page 28 - Size: 88x42x30mm

RS232 Serial InterfaceChapter 10 Peripherals 113FSpO2 Monitor If the fetal monitor has the Combined Interface Module J-14 fitted, you can inter

Page 29 - (1290C option

RS232 Serial Interface114 Chapter 10 Peripherals DIP Switch Settings The DIP switches for the FSpO2 monitor must be set as follows:DIP switches

Page 30 - (CPJ840J5)

Telemetry SystemChapter 10 Peripherals 115Telemetry SystemThe following table lists the signals at each pin. There is no external parameter inp

Page 31 - Installing the Monitor

Telemetry System116 Chapter 10 Peripherals Key: (L) means Logic input is low (0 Volts).16 Ext Par Input17 Gnd Analog Signal Ground.18 Gnd19

Page 32 - Fitting the Monitor to a Wall

80235A (OBMS), M1370A (ODIS), and OB TraceVueChapter 10 Peripherals 11780235A (OBMS), M1370A (ODIS), and OB TraceVueTable 10-6 lists the signal

Page 33

80235A (OBMS), M1370A (ODIS), and OB TraceVue118 Chapter 10 Peripherals Key (L)\means Logic input/output is low (0 V).(H)\means Logic input is h

Page 34

Chapter 11 Replacing Parts 11911Replacing PartsIntroductionThis chapter identifies the boards and replacement parts, giving their part numbers,

Page 35 - Configuring the Monitor

Initial InspectionChapter 1 General Information 3Initial InspectionThe monitor and any supporting options ordered are supplied packed in protec

Page 36 - 0 or 1)

Service Tools120 Chapter 11 Replacing Parts Service ToolsThe basic tools you will need to service the Series 50 A and Series 50 IP-2 fetal monit

Page 37 - Menu Setting Options Default

Lists of PartsChapter 11 Replacing Parts 121Table 11-1 Boards Parts ListItem Description Part Number1 M1351A Single Front End Board rectangular

Page 38 - C03 is displayed

Lists of Parts122 Chapter 11 Replacing Parts Monitor Parts Figure 11-2Monitor Parts Diagram (1 of 3)Figure 11-3Monitor Parts Diagram (2 of 3)

Page 39

Lists of PartsChapter 11 Replacing Parts 123Figure 11-4Monitor Parts Diagram (3 of 3)Table 11-2 lists currently available monitor parts. If a p

Page 40 - Service

Lists of Parts124 Chapter 11 Replacing Parts 11 Chassis - M1353-00101 112 Insulator - M1353-45401 116 Foot repair kit (both colors) M1353-67901

Page 41 - A: and enter:

Lists of PartsChapter 11 Replacing Parts 125Recorder Parts The following diagram shows you how to assemble the recorder using the older and new

Page 42

Lists of Parts126 Chapter 11 Replacing Parts Monitor Housing ColorTable 11-4 lists currently available miscellaneous parts. If a part is still a

Page 43 - Enter. The program will

Lists of PartsChapter 11 Replacing Parts 127Toco Transducer (B lu e , M 1 3 55 A ) Figure 11-6Toco Transducer Note— Blue transducers ar

Page 44 - Enter. The program

Lists of Parts128 Chapter 11 Replacing Parts Ultrasound Transducer (Blue, M1356A) Figure 11-7 Ultrasound Transducer M1356A (Blue) Note— B

Page 45

Lists of PartsChapter 11 Replacing Parts 129DECG Transducer (M1357A) Figure 11-8DECG Transducer M1357A Table 11-7 DECG Transducer: Par

Page 46 - Reading the Options

Overview4 Chapter 1 General Information Overview This guide describes three different Series 50 Fetal Monitor models:M1351A Single Ultrasound mo

Page 47

Lists of Parts130 Chapter 11 Replacing Parts MECG Transducer (M1359A) Figure 11-9MECG Transducer M1359A Table 11-9 MECG Transducer: Part

Page 48

Toco Transducer (Brown)Chapter 11 Replacing Parts 131Toco Transducer (Brown) Figure 11-10Toco Transducer (Brown) Table 11-11Toco Tran

Page 49 - FIC section is additional

Toco Transducer (Brown)132 Chapter 11 Replacing Parts CautionDO NOT use a screwdriver or similar flat-bladed instrument to separate the transdu

Page 50 - Y key to clear the error log

Ultrasound Transducer (Brown)Chapter 11 Replacing Parts 133Ultrasound Transducer (Brown) Table 11-13Toco Transducer: Parts ListItem Description

Page 51

Ultrasound Transducer (Brown)134 Chapter 11 Replacing Parts CautionDO NOT use a screwdriver or similar flat-bladed instrument to separate the tr

Page 52 - Write Options

Patient Module (M1364A)Chapter 11 Replacing Parts 135Patient Module (M1364A) Parts ListTroubleshooting The only functional checks that can be c

Page 53 - FHR Alerting Enable/Disable

Fuses136 Chapter 11 Replacing Parts FusesThe fuse values are printed beside the mains socket: To replace the fuses: 1. Switch off the monito

Page 54 - Error Log Messages

BatteriesChapter 11 Replacing Parts 137BatteriesIf the batteries are not replaced when necessary, the time and date return to their default val

Page 55 - Table 4-5Error Log Messages

Top Cover138 Chapter 11 Replacing Parts Top CoverTo remove the top cover: Figure 11-11Removing the Top Cover 1. Switch off the Monitor and disco

Page 56

Front End BoardChapter 11 Replacing Parts 139Front End Board To remove the board:The illustration above shows removal of a rectangular shaped

Page 57

Major Keys and Parts at a GlanceChapter 1 General Information 5Major Keys and Parts at a GlanceFigure 1-1 General Layout of the Series 50 A and

Page 58

Power Supply Board140 Chapter 11 Replacing Parts Power Supply Board To remove the power supply board:1. Remove the top cover. 2. Disconnect

Page 59 - Theory of Operation

CPU BoardChapter 11 Replacing Parts 141CPU Board To remove the CPU board:1. Remove the top cover. 2. Disconnect the cables (B, C, D, E and F).

Page 60 - Booting and Self Tests

Combined Interface Module142 Chapter 11 Replacing Parts Combined Interface Module To remove the interface module: 1. Remove the retaining screw

Page 61

ChassisChapter 11 Replacing Parts 143Chassis To remove the chassis: 1. Remove the top cover. 2. Remove the combined interface board - if insta

Page 62

On/Off Switch144 Chapter 11 Replacing Parts On/Off Switch To remove the monitor on/off switch:1. Remove the chassis. 2. Remove the two screws

Page 63

Recorder BoardChapter 11 Replacing Parts 145Recorder Board To remove the recorder board: Remove the top cover. 1. Disconnect the cables G, H,

Page 64 - ≤ 3720G

Display Board146 Chapter 11 Replacing Parts Display Board To remove the display board:1. Remove the top cover. 2. Disconnect the cable (H). 3.

Page 65 - ±12V, +5V and +24V

Switch BoardChapter 11 Replacing Parts 147Switch Board To remove the switch board:1. Remove the top cover. 2. Disconnect the cable (L). 3. Pri

Page 66

Loudspeaker148 Chapter 11 Replacing Parts Loudspeaker To remove the loudspeaker:1. Remove the top cover. 2. Disconnect the two wires from the lo

Page 67

TransformerChapter 11 Replacing Parts 149Transformer To remove the transformer: 1. Remove the top cover. 2. Disconnect the three wires from t

Page 68 - Display Board (M1350-66520)

Major Keys and Parts at a Glance6 Chapter 1 General Information 1. Mains socket2. Monitor on/off switch3. Equipotential grounding point4. Monito

Page 69

Drawer Assembly150 Chapter 11 Replacing Parts Drawer Assembly To remove the drawer assembly:Figure 11-13 Removing the Paper Table1. Remove the

Page 70 - Interface Boards

Thermal PrintheadChapter 11 Replacing Parts 151CautionThe use of recorder paper that is not approved by Philips can damage the Monitor. This ty

Page 71

Thermal Printhead152 Chapter 11 Replacing Parts To remove the thermal printhead:1. Place the recorder chassis so that the stepper motor is on th

Page 72

Recorder Sensing AssemblyChapter 11 Replacing Parts 153and slide it into position.5. Refit the recorder chassis into the top cover. 6. Refit th

Page 73 - Interface Board

Stepper Motor154 Chapter 11 Replacing Parts Stepper Motor To replace the stepper motor:1. Remove the two screws securing the stepper motor to

Page 74

IntroductionAppendix A Modem Interface Module 155AModem Interface ModuleIntroductionThe modem interface module allows fetal trace information t

Page 75 - Tests and Error Messages

Connecting Peripheral Devices156 Appendix A Modem Interface Module Connecting Peripheral DevicesAfter you have fitted the module, connect the fo

Page 76

Connecting Peripheral DevicesAppendix A Modem Interface Module 157Barcode Reader Interface Once the barcode reader is installed, check that it

Page 77 - Figure 7-2 Test Pattern

Connecting Peripheral Devices158 Appendix A Modem Interface Module RS232 Serial Interface This RS232 Serial Interface connection enables the mon

Page 78

Entering, Storing, and Transmitting DataAppendix A Modem Interface Module 159Entering, Storing, and Transmitting DataActions associated with th

Page 79 - Operator Error Messages

Monitor Control and Display PanelChapter 1 General Information 7Monitor Control and Display PanelFigure 1-2 Layout of the Monitor Control and D

Page 80

Entering, Storing, and Transmitting Data160 Appendix A Modem Interface Module Setting the Receiver Phone NumberTo enter the Receiver Phone numbe

Page 81 - Troubleshooting Flowcharts

Entering, Storing, and Transmitting DataAppendix A Modem Interface Module 161Keep Patient Dataafter Transmissionor When ClearingTrace MemoryIf

Page 82

Entering, Storing, and Transmitting Data162 Appendix A Modem Interface Module 3. The decimal dots continue to flash until storage is stopped. Th

Page 83 - Error 500: General Failure

PCMCIA Card ModemAppendix A Modem Interface Module 163If transmission problems occur, the fetal monitor automatically attempts to resend the re

Page 84

PCMCIA Card Modem164 Appendix A Modem Interface Module Modem CompatibilityTo ensure compatibility with local telephone systems, modems must be o

Page 85

List of PartsAppendix A Modem Interface Module 165Programming aModem SetupSequence1. Scan the “Start Modem Setup” barcode.2. Use the barcodes t

Page 86

Troubleshooting and Error Messages166 Appendix A Modem Interface Module 2. Err 77 Telephone transmission failed, accompanied by three tones (dis

Page 87 - Error 502: Power Supply

Troubleshooting and Error MessagesAppendix A Modem Interface Module 167If you are unable to transmit a stored fetal trace, having selected the

Page 88

Modem Setup Barcodes168 Appendix A Modem Interface Module Modem Setup BarcodesThe following are the service barcodes for modem setup. See -164 f

Page 89

Modem Setup BarcodesAppendix A Modem Interface Module 169h'6'r'@'%'*'8'%'i'7's'A'$&

Page 90

Monitor Control and Display Panel8 Chapter 1 General Information 5. US1 Signal Quality Indicator. Indicates the quality of the signal detected b

Page 91

Modem Setup Barcodes170 Appendix A Modem Interface Module

Page 92

Safety InformationAppendix B Safety and Environment 171BSafety and EnvironmentSafety InformationFollow these guidelines for maximum patient saf

Page 93

Environment172 Appendix B Safety and Environment The Monitor is designed to fulfill safety requirements according to IEC 60601-1, UL 544 and CSA

Page 94

SpillageAppendix B Safety and Environment 173SpillageIf liquid enters the monitor, you must cease using it immediately. Contact an authorized e

Page 95

Electromagnetic Compatibility174 Appendix B Safety and Environment Avoiding Interference When electromagnetic interference (EMI) is encountered

Page 96

Appendix C Upgrade Key 175CUpgrade KeyIntroductionThis section tells you how to use an upgrade key to upgrade the software of the Philips Serie

Page 97

Upgrade Procedure176 Appendix C Upgrade Key 6. Stick the feature label to the front of the monitor, as in the example below:A: Fetal Movement Pr

Page 98

IndexIndex 177AAccessoriesoptional, 11accessories, 9parts list, 125Bbarcode booklets, part numbers, 10barcode reader, 109connecting, 108fitting holde

Page 99

178 Indexmaternal measurements on the FHR trace, 112maternal parameters, 107baudrate settings, 111NIBP monitors, 111MECG parameter test, 89MECG transd

Page 100 -

Index 179safety test connectorMetron, 101Safety test S2protective earth test, 104Safety test S3patient leakage test, 104Safety test S6enclosure leakag

Page 101

AccessoriesChapter 1 General Information 9AccessoriesSeries 50 A (M1351A)The following accessories are supplied as standard with the Monitor:

Page 102

NoticePhilips makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and

Page 103 - Compatible Monitor

Accessories10 Chapter 1 General Information 8120-2961 (USA).8120-4808 (Europe).One pack of paper:M1910A (USA/Canada)M1911A (Europe)M1913A (Japan

Page 104

AccessoriesChapter 1 General Information 11Options The following accessories can also be supplied when the appropriate option is ordered.Acces

Page 105 - Care and Cleaning

Accessories12 Chapter 1 General Information Accessories Option ModelService and Installation Guide 0B3 Series 50 A and Series 50 IPInstallation

Page 106 - Regular Maintenance

Chapter 2 Technical Specifications 132Technical SpecificationsMonitorPower RequirementsThe monitor is set for the correct voltage at the factor

Page 107

Monitor14 Chapter 2 Technical Specifications DisplaysNumerical Display M1351A Single Ultrasound Model: One heart rate display (orange) and one u

Page 108 - Testing

MonitorChapter 2 Technical Specifications 15Ultrasound Mode System: Pulsed Doppler oscillation.Frequency: 998.4 kHz.Repetition Rate: 3.2 kHz.Ul

Page 109 - (M1364A)

Monitor16 Chapter 2 Technical Specifications Self-Test FacilitiesSelf-test facilities include: System test: With no transducers connected

Page 110 - Safety Testing

Transducers and CablesChapter 2 Technical Specifications 17Transducers and CablesTransducers can be stored at temperatures between -40 and +60°

Page 111

Transducers and Cables18 Chapter 2 Technical Specifications DECG Transducer (M1357A)Input Impedance: >10MΩ (differential, dc to 50/60Hz).

Page 112 - Perform Safety

Transducers and CablesChapter 2 Technical Specifications 19DECG Cable(M1362A)Cable weight: 22 gramsCable Length: 666mm +/- 30mmDECG AdaptorCabl

Page 113

Printing HistoryNew editions are complete revisions of the manual. Update packages, which are issued between editions, contain additional and replacem

Page 114 - Safety Tests

Transducers and Cables20 Chapter 2 Technical Specifications Width: 37mm/1.5in.Height: 21.7mm/0.9in.Operating Temp: 15 to 40°C.IUP Pressure Trans

Page 115 - Instrument Safety

Chapter 3 Installing the Monitor 213Installing the MonitorFitting the Monitor to a SurfaceThe monitor can be mounted on (but not fixed to) an e

Page 116

Fitting the Monitor to a Wall22 Chapter 3 Installing the Monitor Fitting the Monitor to a WallBefore you can fit the monitor to the wall, you mu

Page 117 - Peripherals

Fitting the Paper Take-Up TrayChapter 3 Installing the Monitor 23Fitting the Paper Take-Up TrayTo fit the paper take-up tray to the monitor:

Page 118 - Connecting Peripheral Devices

Carts24 Chapter 3 Installing the Monitor CartsYou can mount the Series 50 A and the Series 50 IP-2 fetal monitors on the CL, CM or CX carts. Cle

Page 119 - Pin Signal Details

Chapter 4 Configuring the Monitor 254Configuring the MonitorIntroductionThis chapter describes how to configure the monitor including time and

Page 120 - RS232 Serial Interface

Configuring the Monitor Using Pushbuttons26 Chapter 4 Configuring the Monitor Configuring the Monitor Using PushbuttonsYou can use the pushbutto

Page 121

Configuring the Monitor Using PushbuttonsChapter 4 Configuring the Monitor 27 Table 4-2Configuration OptionsMenu Set

Page 122 - Measurements

Configuring the Monitor Using Pushbuttons28 Chapter 4 Configuring the Monitor Note— After you put paper into the recorder, let it run for a few

Page 123 - 3cm/min 1 minute

Configuring the Monitor Using BarcodesChapter 4 Configuring the Monitor 29Configuring the Monitor Using BarcodesThe time format, date format, I

Page 125 - Telemetry System

Configuring the Monitor Using a PC30 Chapter 4 Configuring the Monitor Installing the Service ProgramBefore you install the program, ensure that

Page 126

Configuring the Monitor Using a PCChapter 4 Configuring the Monitor 31Connecting the PC to the MonitorFigure 4-2 Service SocketConnect the cabl

Page 127

Configuring the Monitor Using a PC32 Chapter 4 Configuring the Monitor Main MenuTo select an item from the menu, move the cursor to the item you

Page 128 - Ω, 100nF

Configuring the Monitor Using a PCChapter 4 Configuring the Monitor 331. Using the appropriate keys, move the cursor to the selection you requi

Page 129 - Replacing Parts

Configuring the Monitor Using a PC34 Chapter 4 Configuring the Monitor Adjusting theRecorderThe Recorder Heat-Adjust is to compensate for differ

Page 130 - Lists of Parts

Configuring the Monitor Using a PCChapter 4 Configuring the Monitor 35Service Menu To select an item from the menu, move the cursor to the

Page 131 - Item Description Part Number

Configuring the Monitor Using a PC36 Chapter 4 Configuring the Monitor Reading the OptionsThis lists the options installed on the monitor. Press

Page 132 - Monitor Parts

Configuring the Monitor Using a PCChapter 4 Configuring the Monitor 37Reading the Error LogThis displays the error log. (See -44 for the list o

Page 133

Configuring the Monitor Using a PC38 Chapter 4 Configuring the Monitor Printing the Error LogTo print the error log, complete the following step

Page 134 - Enhanced version

Configuring the Monitor Using a PCChapter 4 Configuring the Monitor 39The error log lists the following (above the line and left to right):Erro

Page 135

Contents v1. General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Int

Page 136 - ≥ 3817G02995

Configuring the Monitor Using a PC40 Chapter 4 Configuring the Monitor Clearing the Error LogTo clear the error log, complete the following step

Page 137 - Transducer

Configuring the Monitor Using a PCChapter 4 Configuring the Monitor 41Writing/Resetting the Serial Number and Setting the OptionsNote— You can

Page 138 - (Blue, M1356A)

Configuring the Monitor Using a PC42 Chapter 4 Configuring the Monitor To select an option:1. Using the appropriate keys, move the cursor to the

Page 139 - (M1357A)

Configuring the Monitor Using a PCChapter 4 Configuring the Monitor 43FHR Alerting Enable/DisableThis screen enables you to enable or disable t

Page 140 - (M1359A)

Error Log Messages44 Chapter 4 Configuring the Monitor Error Log MessagesTable 4-5Error Log MessagesMessage Location of Error and SolutionErr 18

Page 141

Error Log MessagesChapter 4 Configuring the Monitor 45Err 611 Wrong loudspeaker impedance. See the Troubleshooting Flowchart in Chapter 8.Err

Page 142 - Ω p/n: 0698-4443

Error Log Messages46 Chapter 4 Configuring the Monitor

Page 143 - Ultrasound Transducer (Brown)

Chapter 5 Setting Time, Date, and Paper Speed 475Setting Time, Date, and Paper SpeedTime and DateTo set the time and date: Figure 5-1

Page 144 - Ω p/n: 0698−3511

Paper Speed48 Chapter 5 Setting Time, Date, and Paper Speed Paper SpeedTo set the paper speed: Figure 5-2 Paper Speed Controls1. Press

Page 145 - Patient Module (M1364A)

System OverviewChapter 6 Theory of Operation 496Theory of OperationSystem OverviewThe boards and their interconnections are shown in the follow

Page 146

vi ContentsFitting the Monitor to the Angle Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 147 - Batteries

Booting and Self Tests50 Chapter 6 Theory of Operation Booting and Self Tests The monitor carries out a basic test every time it is switched on,

Page 148 - Top Cover

Booting and Self TestsChapter 6 Theory of Operation 51The Combined Interface Board and the Modem Interface Board both have processors that test

Page 149 - Front End Board

Front End Board52 Chapter 6 Theory of Operation Front End BoardM1351A (M1353-66501 and M1353-66511) The Frontend board (M1353-66501 − single,

Page 150 - Power Supply Board

Front End BoardChapter 6 Theory of Operation 53controls the ECG. Each type of transducer has a specific mode resistor which is recognized by th

Page 151 - CPU Board

Front End Board54 Chapter 6 Theory of Operation Frontend Board for M1353A (M1350-66517) Used in monitors with serial number prefix ≤ 3720G.Figur

Page 152 - Modem Interface Module

Power Supply Board (M1353-66502)Chapter 6 Theory of Operation 55Power Supply Board (M1353-66502) The Power Supply Board operates as a linear re

Page 153 - Chassis

CPU Board (M1353-66503 and M1353-66513)56 Chapter 6 Theory of Operation CPU Board (M1353-66503 and M1353-66513) CPU Board M1353-66503 This board

Page 154 - On/Off Switch

CPU Board (M1353-66503 and M1353-66513)Chapter 6 Theory of Operation 57CMOS RAM memory, backed by a battery to store short time device settings

Page 155 - Recorder Board

Display Board (M1350-66520)58 Chapter 6 Theory of Operation Display Board (M1350-66520) The display board consist of the following main parts:Ke

Page 156 - Display Board

Recorder Interface Board (M1353-66510)Chapter 6 Theory of Operation 59Recorder Interface Board (M1353-66510) The Recorder Board provides the in

Page 157 - Switch Board

Contents viiIntroduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 158 - Loudspeaker

Interface Boards60 Chapter 6 Theory of Operation Interface Boards The interface boards available are:The Combined Interface BoardThe Modem In

Page 159 - Transformer

Interface BoardsChapter 6 Theory of Operation 61Figure 6-3 Combined Interface Board M1353-66531(E)Both the M1353-66531/66531E boards include t

Page 160 - Drawer Assembly

Interface Boards62 Chapter 6 Theory of Operation System Section:This section is controlled by an 80C51 microcontroller. The communication betwee

Page 161 - Thermal Printhead

Interface BoardsChapter 6 Theory of Operation 63Modem Interface BoardThe M1353-66532 modem interface board allows the storage and the transmiss

Page 162

Interface Boards64 Chapter 6 Theory of Operation

Page 163 - Recorder Sensing Assembly

Chapter 7 Tests and Error Messages 657Tests and Error MessagesService PhilosophyAny errors produced are marked on the trace paper with the symb

Page 164 - Stepper Motor

Tests and Error Messages66 Chapter 7 Tests and Error Messages Err xxx , time and date are printed on the paper after 13 seconds, and then every

Page 165

Tests and Error MessagesChapter 7 Tests and Error Messages 67Figure 7-2 Test Pattern4. Check the dots in the test pattern to ensure all the hea

Page 166

Tests and Error Messages68 Chapter 7 Tests and Error Messages Parameter Test This tests the signal to and from the transducer, but not the trans

Page 167 - Interface

Tests and Error MessagesChapter 7 Tests and Error Messages 69Operator Error Messages Table 7-2 Operator Error MessagesMessage Display Cause and

Page 168 - RS232 Serial

viii ContentsOrdering Parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 169

Tests and Error Messages70 Chapter 7 Tests and Error Messages Permanent TestNote This is similar to the Cyclic Test described on page 35. To car

Page 170

Chapter 8 Troubleshooting Flowcharts 718 Troubleshooting FlowchartsIntroductionTroubleshooting is based on the interpretation of the error mess

Page 171 - Clearing

Error 500: General Failure72 Chapter 8 Troubleshooting Flowcharts Error 500: General FailureWarningThe following task requires that the power be

Page 172

Error 500: General FailureChapter 8 Troubleshooting Flowcharts 73Figure 8-2 Connectors and LEDs

Page 173 - PCMCIA Card Modem

Error 501, 511, 512, 516, 517: Front End Board74 Chapter 8 Troubleshooting Flowcharts Error 501, 511, 512, 516, 517: Front End BoardWarningThe f

Page 174 - TraceVue in the

Error 501, 511, 512, 516, 517: Front End BoardChapter 8 Troubleshooting Flowcharts 75Figure 8-4 Connectors and LEDs

Page 175 - List of Parts

Error 502: Power Supply76 Chapter 8 Troubleshooting Flowcharts Error 502: Power SupplyWarningThe following task requires that the power be on wh

Page 176 - Connection

Error 502: Power SupplyChapter 8 Troubleshooting Flowcharts 77Figure 8-6 Connectors and FusesSCREW

Page 177

Error 503 and 513: CPU Board78 Chapter 8 Troubleshooting Flowcharts Error 503 and 513: CPU Board Figure 8-7 Troubleshooting: Error 503

Page 178 - Modem Setup Barcodes

Error 510: Recorder BoardChapter 8 Troubleshooting Flowcharts 79Error 510: Recorder Board Figure 8-8 Troubleshooting: Error 510See Fi

Page 179

Contents ixModem Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 180

Error 531: Combined Interface Board80 Chapter 8 Troubleshooting Flowcharts Error 531: Combined Interface Board Figure 8-9 Troubleshoot

Page 181 - Safety and Environment

Error 532: Modem Interface BoardChapter 8 Troubleshooting Flowcharts 81Error 532: Modem Interface Board Figure 8-10 Troubleshooting:

Page 182 - Environment

Error 70: Modem Not Responding82 Chapter 8 Troubleshooting Flowcharts Error 70: Modem Not Responding Figure 8-11 Troubleshooting: Er

Page 183 - Spillage

Error 77: Modem Transmission FailureChapter 8 Troubleshooting Flowcharts 83Error 77: Modem Transmission Failure Figure 8-12 Troubl

Page 184 - Interference

Error 601: Paper Feed84 Chapter 8 Troubleshooting Flowcharts Error 601: Paper Feed Figure 8-13 Troubleshooting: Error 601See Figure

Page 185 - Upgrade Key

Error 610: No LoudspeakerChapter 8 Troubleshooting Flowcharts 85Error 610: No Loudspeaker Figure 8-14 Troubleshooting: Error 610

Page 186 - Upgrade Procedure

Error 611: Loudspeaker86 Chapter 8 Troubleshooting Flowcharts Error 611: Loudspeaker Figure 8-15 Troubleshooting: Error 611

Page 187 - Index 177

Ultrasound Parameter TestChapter 8 Troubleshooting Flowcharts 87Ultrasound Parameter Test Figure 8-16 Troubleshooting: Ultrasound Par

Page 188 - 178 Index

DECG Parameter Test88 Chapter 8 Troubleshooting Flowcharts DECG Parameter Test Figure 8-17 Troubleshooting: Direct ECG Parameter Testre

Page 189 - Index 179

MECG Parameter TestChapter 8 Troubleshooting Flowcharts 89MECG Parameter Test Figure 8-18 Troubleshooting: Maternal ECG Parameter Tes

Commentaires sur ces manuels

Pas de commentaire