Measurement Specifications 34 Installation and Specifications
355
Temp
Invasive Pressure Alarm Specifications Range Adjustment Delay
Pressure –40 to 360 mmHg
(–5.0 to 48 kPa)
-40 to 30 mmHg
2 mmHg (0.5 kPa)
> 30 mmHg
5 mmHg (1 kPa)
max. 12
seconds
Pulse 25 to 300 bpm Adult:
1 bpm steps (25 to 40 bpm)
5 bpm steps (40 to 300 bpm)
Pedi/Neo:
1 bpm steps (25 to 50 bpm)
5 bpm steps (50 to 300 bpm)
Tachycardia Difference to high limit 0 to
50 bpm
5 bpm steps max. 14
seconds
Clamping at 150 to 300
bpm
5 bpm steps
Bradycardia Difference to low limit 0 to
50 bpm
5 bpm steps max. 14
seconds
Clamping at 25 to 100 bpm
5 bpm steps
Temp Performance Specifications
Temp Range –1 to 45 °C (30 to 113 °F)
Resolution
0.1 °C (32.2 °F)
Accuracy
±0.1 °C (±0.2 °F)
Average Time Constant Less than 10 seconds
Alarms Range –1 to 45 °C (30 to 113 °F)
Adjustment
-1 to 35 °C (30 to 95 °F): 0.5 °C (1.0 °F) steps
35 to 45 °C (95 to 113 °F): 0.1 °C (0.2 °F) steps
Temp Alarm Specifications Range Adjustment
Tem p Hi gh/ Low Al a rms
–1 to 45 ºC (30 to 113 ºF) -1 to 35 °C (30 to 95 °F), 0.5 °C (1.0 °F) steps
35 to 45 °C (95 to 113 °F), 0.1 °C (0.2 °F) steps