Cardiac Output Default Settings 35 Default Settings Appendix
377
Cardiac Output Default Settings
Mean Only No Yes
Filter 12 Hz 12 Hz
Mercury Cal. Yes Yes
Artifact Suppr. 60 sec 60 sec
Unit mmHg mmHg
Color Yellow Magenta
Invasive Pressure
Settings
PAP Settings ICP, IC1, IC2 Settings
Factory Adult Factory Pedi Factory Neo Factory Adult Factory Pedi Factory Neo
C.O. Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
Auto-Calibration On
Tblood High Limit 39.0 ºC
Tblood Low Limit 36.0 ºC
Tblood Alarm On/Off On
Tem per atu re Un it
0
C
Color Green Yellow
CCO/CCI Settings Factory Adult Factory Pedi Factory Neo
Settings common to CCO and CCI
Alarms from CCO
CCO from ABP
Color Green
CCO
Alarms On
CCO High Limit 8.5 l/min 3.7 l/min 1.3 l/min
CCO Low Limit 4.0 l/min 2.6 l/min 0.3 l/min
CCI
Alarms On
CCI High Limit 4.3 l/min/m
2
3.7 l/min/m
2
5.2 l/min/m
2
CCI Low Limit 2.0 l/min/m
2
2.6 l/min/m
2
1.2 l/min/m
2