Philips V26CT Manuel d'utilisateur Page 210

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 392
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 209
oxyCRG Recordings
6-38 Recording Functions
Recording Functions
oxyCRG
Alarm
Recording
Pressing the softkey allows you to choose
between “Standard” alarm recording or “oxyCRG” alarm recording.
If you select “oxyCRG”, an oxyCRG recording is started when an alarm
occurs. Six minutes of pre-alarm data and two minutes of post-alarm
data are recorded. Any manually initiated recording that is in progress
will be interrupted, and the oxyCRG recorded instead. If a second alarm
occurs while an oxyCRG alarm recording is already in progress, the
recording will be extended to two minutes after the second alarm
appears on the strip.
If you select “Standard”, your selected realtime waves are recorded
when an alarm occurs, as described earlier in this chapter.
Monitor Setup
Recordng Setup
Delayed & Alarm
Change AlRecTyp
Vue de la page 209
1 2 ... 205 206 207 208 209 210 211 212 213 214 215 ... 391 392

Commentaires sur ces manuels

Pas de commentaire