Controls and Connectors
Monitor Installation and Patient Safety 11-25
Monitor Installation
and Patient Safety
M1094B Display Module
Controls of the
M1092A,
M1094A and
M1094B
Displays
1. Video Termination Switch; thus switch is used to terminate the
video signal at this display: this must be switched to the “75
position if only a master display is used and must be switched to the
“HIGH IMP.” position if a second, duplicate display is used.
Maximum Voltages
System Power Connector 60 V
Human Interface Connector 12 V
Local Power Connector 240 V
Video Out Connector 5 V
Video In Connector 5 V